MRL

Materials Research Laboratory

About    Directory    Contact Us    For Visitors    Search MRL    Text Only

CMM- Transmission and Scanning Transmission Electron Microscopy (TEM and STEM)

Instrumentation for Transmission and Scanning Transmission Electron Microscopy:

TEM and STEM

Staff Contact:



Wacek Swiech
(217) 244-9532
wswiech@illinois.edu

Mike Marshall
(217) 265-5380
mtmarsha@illinois.edu

JEOL 2100 Cryo TEM

Usage Rules
Links

Location
0011 Materials Research Laboratory

General:
The 2100 Cryo is a conventional TEM with large pole piece gap. It is optimized for BF/DF imaging, diffraction and high sample tilts. It operates at 200kV. The 2100 Cryo is equipped with a combined video rate/slow scan camera for real time imaging and final image recording. It has a low-dose mode. A high tilt holder is available with +/-60 degrees of tilt. Cryo transfer stage, double tilt heating, and liquid N2 stages are available.

Specifications:

  • Cs = 2.0mm
  • Energy 200kV
  • Tilt 60 degrees on two axes
  • Point resolution = 0.27nm
  • Information Limit > 0.14nm
  • Lens Type-Side entry (Condenser/Objective)

Attachments:

  • Digital Imaging - Gatan MatScan1kx1k progressive scan CCD
  • Computer Control - Digital Micrograph
  • Free Lens Control
  • Cryo transfer stage
  • 1000ºC double tilt holder
  • Liquid N2 double tilt holder